FAQs

Fluidics FAQs

Here we have provided comprehensive lists of commonly asked questions regarding our fluidics products and related applications. This information is designed to support your inquiries, but if you don’t find the answers you are looking for we encourage you to contact us for further assistance.

Please use the sort buttons in the left navigation to navigate between questions and answers that are specific to our Optical Filters, Optical Imaging Systems, Cameras, Microfluidics, Webstore, and Engineering Partnership.

I’m looking for a simple biocompatible Semi-Prep Inline filter; what do you have available?

IDEX Health & Science has offers a high-pressure, biocompatible Semi-Prep Inline Filter. It consists of a stainless steel body, two PEEK end fittings and a replaceable PEEK frit. Once you buy the unit, the only part that will need replacing is the PEEK frit inside. Available with either 2µm or 10µm frits. The 2µm version has a pressure drop of approximately 40-45 psi at a flow rate of 100 mL/min of water across the frit. The 10µm version has a pressure drop of less then 10 psi under the same conditions.

Please Note: our Semi-Prep Inline Filters are engineered to be used with 1/16″ OD tubing utilizing polymer fittings, such as our one-piece F-120 Fingertight fitting.