FlushNut™ Super Flangeless™ Fittings Instructions

by IDEX Health & Science

FlushNut™ Super Flangeless™ Fittings from IDEX Health & Science

  1. Slide the FlushNut, Stainless Steel Compression Ring and Super Flangeless Ferrule, in that order, onto the tubing ensuring the tubing end is flush with the base of the ferrule. The side of the ring with the wider flat surface should face toward the nut.
  2. Insert the assembly into the receiving port. Hold the tubing against the bottom of the port while finger tightening the nut into place.

Please Note: IDEX Health & Science highly recommends using the FlushNut Wrench for installing and removing FlushNuts. Using the FlushNut Wrench, while continuing to hold the tubing down into the receiving port, tighten the FlushNut until the ferrule is completely seated.