Recommended Fittings for 25 μm ID Tubing
by
IDEX Health & Science
| Tube OD | Port | Port Material | Nut | Ferrule | Sleeve | Torque | PSI Rating |
| .062” | 10-32 coned | SS | F-196 | F-192 | N/A | 3 inch / pounds | 7000 |
| .062” | 10-32 coned | PEEK | F-196 | F-192 | N/A | 4 inch / pounds | 6500 |
| .032” | 10-32 coned | PEEK | F-331N | STD | F-247 | 4 inch / pounds | 4000 |
| .032” | 6-32 coned | PEEK | F-126H | N/A | N/A | 2 inch / pounds | 7000 |
| 360 μm | 10-32 coned | PEEK | F-331N | STD | F-242 | 4 inch / pounds | 4000 |
| 360 μm | 6-32 coned | PEEK | F-124H | N/A | N/A | 2 inch / pounds | 7000 |
Instructions for Use
- 1/16” OD tubing should be scored and snapped with the IDEX Health & Science tubing cutter (A-327). Twirl cutting may cause occlusion of 25 μm ID.
- 1/32” and 360 μm tubing should be cut with the IDEX Health & Science tubing cutter (A-350)
- Tightening more than published torque may occlude tubing ID.