This set is designed to take advantage of the superior resolution, sensitivity, and image fidelity of today's state-of-the-art laser-based microscopes. The filter spectra are optimized to capture the maximum fluorescence from Cy5 and other similar fluorophores while offering superb blocking of popular 635 nm (solid-state laser diode), 633 nm (HeNe gas) or 647 nm (Kr-ion gas) lasers. The dichroic offers industry-leading 1λ P-V RWE for minimal focus shift and aberrations of the laser beam spot up to 10 mm diameter to enable popular imaging and Super-Resolution techniques including TIRF, PALM, STORM, and Structured-Illumination.
Loose and cube mounted filter sets are certified to deliver dichroic flatness <= λ/2 PV per inch @ 632.8 nm.
FF01-640/14-25FL-004697
Each
|
FL-004697 |
|
BLP01-635R-25FL-008533
Each
|
FL-008533 |
|
|
S-000995 |
FL-004697 / 0
FL-006959 / 0
FL-008533 / 0
1 of 1
This set is mounted in an Evident Scientific (Olympus) OFF Fluorescence Filter Holder, which is compatible with the BX3 (BX53, BX63) and IX3 Series (IX53, IX73, IX83) of Evident Scientific (Olympus) microscopes.
Filter Role | Filter Size | Maximum Thickness |
Exciter (Diameter) | 25.0 mm | 5.0 mm |
Emitter (Diameter) | 25.0 mm | 5.0 mm |
Dichroic (L × W) | 26 mm x 38 mm | 1.5 mm |
Topic | Description |
Cleaning Optical Filters | Instructions for cleaning hard coated optical filters |
Filter Reliability | Compares the performance and reliability of hard coated filters to traditional soft-coated filters and provides information about the testing standards for Semrock filters |
Working with Optical Density | Tables to describe the relationship between OD and transmission with tips for quickly calculating OD |
What is Pixel Shift, and how does the Semrock ZERO™ option help? | Why Pixel Shift degrades multicolor fluorescence imaging performance, and how the ZERO™ option mitigates this for Avant and BrightLine filter sets. |